أجهزة طبية
RotaFlow Console
Summary
| Type | Health product recall |
|---|---|
| Product | RotaFlow Console |
| Issue | Medical devices - Performance |
| What to do | Contact the manufacturer if you require additional information. |
| Audience | Healthcare |
| Brand(s) | MAQUET CARDIOPULMONARY GMBH |
| Last updated | 2026-10-29 |
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
RotaFlow Console |
More than 10 numbers, contact manufacturer. |
70104.3292 |
Issue
During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.
Details
| Original published date | 2026-10-29 | ||
|---|---|---|---|
| Alert / recall type | Health product recall | ||
| Category | Health products - Medical devices - Cardiovascular | ||
| Companies |
| ||
| Published by | Health Canada | ||
| Audience | Healthcare | ||
| Recall class | Type II | ||
| Recall date | 2026-10-27 | ||
| Identification number | RA-82878 |