Philips Azurion System

تفاصيل الاستدعاء أدناه منشورة بالإنجليزية كما أصدرتها الحكومة الكندية دون تعديل. لم تنشر الحكومة صوراً لهذا المنتج.

Summary

TypeHealth product recall
ProductPhilips Azurion System
IssueMedical devices - Performance
What to do

Contact the manufacturer if you require additional information.

AudienceHealthcare
Brand(s)PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Last updated2026-10-29

Affected products

Affected products

Lot or serial number

Model or catalogue number

Azurion 7 M20

Not applicable.

722 224

Azurion 7 M20

Not applicable.

722 079

Azurion 7 B12

Not applicable.

722 067

Azurion 7 M12

Not applicable.

722 223

Azurion 7 M12

Not applicable.

722 078

Azurion 7 B20

Not applicable.

722 226

Azurion 7 B20

Not applicable.

722 068

Issue

Philips has identified a safety issue affecting certain Azurion systems potentially leading to loss of imaging functionality. 
The power distribution unit is an internal assembly of the Philips Azurion system that contains multiple power supplies. Philips has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system.
When a power supply failure occurs during a start or restart, the power distribution unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use.

Details

Original published date2026-10-29
Alert / recall typeHealth product recall
CategoryHealth products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands, 5684 PC
Published byHealth Canada
AudienceHealthcare
Recall classType II
Recall date2026-10-23
Identification numberRA-82876