أجهزة طبية
Magnetom Free.Max
Summary
| Type | Health product recall |
|---|---|
| Product | Magnetom Free.Max |
| Issue | Medical devices - Performance |
| What to do | Contact the manufacturer if you require additional information. |
| Audience | Healthcare |
| Brand(s) | Siemens Shenzhen Magnetic Resonance Ltd. |
| Last updated | 2026-10-27 |
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Magnetom Free.Max |
206217 |
11408425 |
Magnetom Free.Max |
206217 |
11408424 |
Issue
The service pack NX VA60A-SP04 is a bug fix update (full installation) that solves/improves the following:
- Latest it security deliveries and increase in overall stability (syngo MR XA60A-SD12 content included)
- Installation of multiple SW delta packages now possible at the same time
- EU DataAact (EY) 2023/2854 implementation:access to all machine data
- Solved image loading errors for MR Breast Biopsy
- General MR Protocol Manager improvements
- Coil specifications adjusted for BM Body 18 Pilot Tone and Spine 32 and 1.5T Sentinelle Breast coils image brightness
- Supporting IEC 60601-1 compliance, enabling predictable and safe system behavior with regard to table movement
Details
| Original published date | 2026-10-27 | ||
|---|---|---|---|
| Alert / recall type | Health product recall | ||
| Category | Health products - Medical devices - Radiology | ||
| Companies |
| ||
| Published by | Health Canada | ||
| Audience | Healthcare | ||
| Recall class | Type II | ||
| Recall date | 2026-10-20 | ||
| Identification number | RA-82867 |
Source: Government of Canada, Recalls and safety alerts – https://recalls-rappels.canada.ca/en/alert-recall/magnetom-freemax