DxC 500i Clinical Analyzer and DxC 500 AU Clinical Chemistry Analyzer
Summary
| Type | Health product recall |
|---|---|
| Product | DxC 500i Clinical Analyzer and DxC 500 AU Clinical Chemistry Analyzer |
| Issue | Medical devices - Performance |
| What to do | Contact the manufacturer if you require additional information. |
| Audience | Healthcare |
| Brand(s) | BECKMAN COULTER, INC. |
| Last updated | 2026-10-26 |
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
DxC 500 AU Clinical Chemistry Analyzer |
1.5 or lower |
C63520 |
DxC 500 AU Clinical Chemistry Analyzer |
1.5 or lower |
C63519 |
DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module |
1.5 or lower |
C63521 |
DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module |
1.5 or lower |
C63522 |
DxC 500i Clinical Analyzer, DxC 500i Access 2 Immunoassay Module |
1.5 or lower |
C13252 |
Issue
Beckman Coulter has determined that a modification made by operators to 25 predefined calculated rules, which are included as part of the parameter package, will revert to the default expression following parameter package or software upgrades.
Details
| Original published date | 2026-10-26 | ||
|---|---|---|---|
| Alert / recall type | Health product recall | ||
| Category | Health products - Medical devices - Chemistry | ||
| Companies |
| ||
| Published by | Health Canada | ||
| Audience | Healthcare | ||
| Recall class | Type II | ||
| Recall date | 2026-10-15 | ||
| Identification number | RA-82863 |
Source: Government of Canada, Recalls and safety alerts – https://recalls-rappels.canada.ca/en/alert-recall/dxc-500i-clinical-analyzer-and-dxc-500-au-clinical-chemistry-analyzer-0