PD-L1 IHC 22C3 pharmDx
Summary
| Type | Health product recall |
|---|---|
| Product | PD-L1 IHC 22C3 pharmDx |
| Issue | Medical devices - Labelling and packaging |
| What to do | Contact the manufacturer if you require additional information. |
| Audience | Healthcare |
| Brand(s) | Agilent Technologies, Inc. |
| Last updated | 2026-10-28 |
Affected products
Affected product |
Lot or serial number |
Model or catalogue number |
|---|---|---|
PD-L1 IHC 22C3 pharmDx |
12001736B |
GE006 |
PD-L1 IHC 22C3 pharmDx |
12010368B |
GE006 |
PD-L1 IHC 22C3 pharmDx |
11992577B |
GE006 |
Issue
Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx (GE006), revision 02 (PD005650). Due to an inadvertent formatting change in section "9.2 tissue sections," tissue-section thickness was incorrectly listed as 45 μm instead of 4-5 μm, and slide-storage temperature was incorrectly listed as 28 °c instead of 2-8 °c. This issue is limited to the electronic IFU (eIFU) associated with the three lots.
Details
| Original published date | 2026-10-28 | ||
|---|---|---|---|
| Alert / recall type | Health product recall | ||
| Category | Health products - Medical devices - Microbiology | ||
| Companies |
| ||
| Published by | Health Canada | ||
| Audience | Healthcare | ||
| Recall class | Type II | ||
| Recall date | 2026-10-15 | ||
| Identification number | RA-82862 |