Summary

TypeHealth product recall
ProductFreeStyle Libre 3 App
IssueMedical devices - Device compatibility
Medical devices - Performance
What to do

For further details on this recall, please refer to the manufacturer's website at: HTTPS://WWW.LIBRECHECK.COM/EN-CA/HOME.HTML

Brand(s)Abbott Diabetes Care Ltd.
Last updated2026-10-26

Affected products

Affected products

Lot or serial number

Model or catalogue number

FreeStyle Libre 3 App (Android)

version 3.6.4

V3.4

FreeStyle Libre 3 App (Android)

version 3.6.7

V3.4

FreeStyle Libre 3 App (Android)

version 3.6.8

V3.4

FreeStyle Libre 3 App (Android)

version 3.6.5

V3.4

FreeStyle Libre 3 App (Android)

version 3.6.6

V3.4

FreeStyle Libre 3 App (iOS)

version 3.6.4

V3.4

FreeStyle Libre 3 App (iOS)

version 3.6.7

V3.4

FreeStyle Libre 3 App (iOS)

version 3.6.8

V3.4

FreeStyle Libre 3 App (iOS)

version 3.6.5

V3.4

FreeStyle Libre 3 App (iOS)

version 3.6.6

V3.4

Issue

Abbott has identified an important security update for the FreeStyle Libre 3 App. This software update is required to maintain security of the apps.  Following the security update to the FreeStyle Libre 3 App (Android and iOS), the software update implemented to mitigate the sensor impersonation vulnerability introduced an unintended defect that, under certain conditions, can result in prolonged loss of communication between the mobile application and the sensor. Customers are advised to immediately update their app to the latest version 3.6.10 to rectify the issues.

Details

Original published date2026-10-26
Alert / recall typeHealth product recall
CategoryHealth products - Medical devices - General hospital and personal use
Companies

Abbott Diabetes Care Ltd.

Range Road, Witney, Oxfordshire, United Kingdom, OX290YL

Published byHealth Canada
AudienceGeneral public
Recall classType II
Recall date2026-10-16
Identification numberRA-82855

Source: Government of Canada, Recalls and safety alerts – https://recalls-rappels.canada.ca/en/alert-recall/freestyle-libre-3-app-0

خدمات أخرى من هلا كندا