أجهزة طبية
HeartSine® Samaritan® PAD 350P/360P/500P
Summary
| Type | Health product recall |
|---|---|
| Product | HeartSine® Samaritan® PAD 350P/360P/500P |
| Issue | Medical devices - Performance |
| What to do | Contact the manufacturer if you require additional information. |
| Brand(s) | Heartsine Technologies, Ltd. |
| Last updated | 2026-11-25 |
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
HeartSine® Samaritan® PAD 350p |
More than 10 numbers, contact manufacturer. |
SAM 350P |
HeartSine® Samaritan® PAD 500p |
More than 10 numbers, contact manufacturer. |
SAM 500P |
HeartSine® Samaritan® PAD 360p |
More than 10 numbers, contact manufacturer. |
SAM 360P |
Issue
Stryker has identified a potential issue with the electrical connection between Pad-Pak ™ or Pediatric-Pak™ and the AED. The connection uses spring-loaded electrical contacts, also known as pogo pins. Over time, oxidation of the pogo pin may increase electrical resistance and weaken the internal spring. If the condition progresses, the electrical connection between the Pad-Pak ™ or Pediatric-Pak™ battery and the AED may become intermittent or may be lost.
Details
| Original published date | 2026-11-25 |
|---|---|
| Alert / recall type | Health product recall |
| Category | Health products - Medical devices - Cardiovascular |
| Companies | Heartsine Technologies, Ltd. 207 Airport Road West, Belfast, Northern Ireland, United Kingdom, BT3 9ED |
| Published by | Health Canada |
| Audience | General public |
| Recall class | Type I |
| Recall date | 2026-11-23 |
| Identification number | RA-83012 |
Source: Government of Canada, Recalls and safety alerts – https://recalls-rappels.canada.ca/en/alert-recall/heartsiner-samaritanr-pad-350p360p500p